[5048792] It was reported that the video cart shut off, allegedly causing 2-3 minutes loss of image while the equipment rebooted.
Patient Sequence No: 1, Text Type: D, B5


[12469739] Additional information will be provided once the investigation has been completed.
Patient Sequence No: 1, Text Type: N, H10


[30345548] The reported device was not received for investigation; therefore, the reported failure cannot be confirmed. The complaint will be closed without a detailed investigation report and based on probable root cause. In the event that the device is received, the complaint will be reopened, a full evaluation will be conducted, and the investigation will be updated with the new results. Probable root cause for the reported failure involving this device could have been caused by: isolation transformer malfunctions. Power strip malfunctions. Circuit overload. Current inrush. Use error. In sum, the reported failure could not be confirmed since the device was not received for investigation.
Patient Sequence No: 1, Text Type: N, H10


[30345549] It was reported that the video cart shut off, allegedly causing 2-3 minutes loss of image while the equipment rebooted.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0002936485-2014-00935
MDR Report Key4256221
Report Source05
Date Received2014-11-17
Date of Report2014-10-20
Date of Event2014-10-20
Date Mfgr Received2014-10-20
Date Added to Maude2014-11-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. THOMAS SHAFER
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal95138
Manufacturer Phone4087542000
Manufacturer G1STRYKER ENDOSCOPY-SAN JOSE
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal Code95138
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTRYKER STANDARD VIDEO CART WITH ISOLATION TRANSFORMER
Generic NameCART, EMERGENCY, CARDIOPULMONARY (EXCLUDING EQUIPMENT)
Product CodeBZN
Date Received2014-11-17
Catalog Number0240099011
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER ENDOSCOPY-SAN JOSE
Manufacturer Address5900 OPTICAL COURT SAN JOSE CA 95138 US 95138


Patients

Patient NumberTreatmentOutcomeDate
10 2014-11-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.