DIMENSION RXL MAX WITH HM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-11-17 for DIMENSION RXL MAX WITH HM manufactured by Siemens Healthcare Diagnostics Inc.

Event Text Entries

[5018911] Discordant, falsely elevated hemoglobin a1c (hb1c) results were obtained on multiple patient samples on a dimension rxl max with hm instrument. The discordant results were reported to the physician(s) who questioned them. The samples were repeated on the same instrument. The corrected hb1c results were reported to the physician(s). There are no known reports of patient intervention or adverse health consequences due to the discordant hb1c patient results.
Patient Sequence No: 1, Text Type: D, B5


[12517533] A siemens customer service engineer (cse) was dispatched to the customer site. The cse evaluated the instrument and instrument data and replaced the reagent 2 and reagent 3 transducer, heat torch and lamp. The cse checked alignments and cleaned windows. The cause of the discordant hb1c results is unknown. The cse successfully ran quality controls. The instrument is performing within specifications. Further evaluation of the device is not required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1226181-2014-00593
MDR Report Key4256883
Report Source05,06
Date Received2014-11-17
Date of Report2014-10-23
Date of Event2014-10-22
Date Mfgr Received2014-10-23
Device Manufacturer Date2007-12-03
Date Added to Maude2014-11-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJOHN NELSON
Manufacturer StreetSIEMENS HEALTHCARE DIAGNOSTICS 511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242530
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC
Manufacturer Street101 SILVERMINE ROAD
Manufacturer CityBROOKFIELD CT 06804
Manufacturer CountryUS
Manufacturer Postal Code06804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDIMENSION RXL MAX WITH HM
Generic NameCLINICAL CHEMSITRY SYSTEM
Product CodePDJ
Date Received2014-11-17
Model NumberDIMENSION RXL MAX WITH HM
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC
Manufacturer Address101 SILVERMINE ROAD BROOKFIELD CT 06804 US 06804


Patients

Patient NumberTreatmentOutcomeDate
10 2014-11-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.