TRANSCUTANEOUS BLOOD GAS MONITOR TCM TOSCA/COMBIM 903-111

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,07,company representative, report with the FDA on 2014-11-15 for TRANSCUTANEOUS BLOOD GAS MONITOR TCM TOSCA/COMBIM 903-111 manufactured by Radiometer Medical Aps.

Event Text Entries

[16182919] The tcm combim was reading too high when monitoring a pt. Readings were 80-90 mmhg and continued to go higher. As a consequence the patient (a pre-term infant) received treatment consisting in "bagging" to bring co2 down. It is stated in the complaint that there was no harm to the pt as result of the treatment. Later a blood sample from the pt was tested for co2 with result of 36 mmhg. At that time the tcm combim was reading 130 mmhg.
Patient Sequence No: 1, Text Type: D, B5


[16385270] A simulator test has been performed with the combim monitor and the test was passed. The sensor electrode has been requested sent back to radiometer for investigation but it has not yet been received.
Patient Sequence No: 1, Text Type: N, H10


[74620945] The sensor electrode was sent back to radiometer for investigation. The result of the investigation was that the sensor electrode was working as intended.
Patient Sequence No: 1, Text Type: N, H10


[74621010] During an internal review of this case, it was discovered that the model and catalog number was stated incorrectly in the initial report and follow-up report no 1.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002807968-2014-00057
MDR Report Key4260898
Report Source05,07,COMPANY REPRESENTATIVE,
Date Received2014-11-15
Date of Report2016-10-14
Date Mfgr Received2016-10-07
Device Manufacturer Date2014-06-01
Date Added to Maude2014-11-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS KRISTINA TOFT
Manufacturer StreetAAKANDEVEJ 21
Manufacturer CityBROENSHOEJ, 2720
Manufacturer CountryDA
Manufacturer Postal2720
Manufacturer Phone538273349
Manufacturer G1RADIOMETER MEDICAL APS
Manufacturer StreetAAKANDEVEJ 21
Manufacturer CityBROENSHOEJ, DK-2700
Manufacturer CountryDA
Manufacturer Postal CodeDK-2700
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRANSCUTANEOUS BLOOD GAS MONITOR
Generic NameTCM TOSCA/COMBIM
Product CodeLKD
Date Received2014-11-15
Returned To Mfg2014-08-01
Model NumberTCM TOSCA/COMBIM
Catalog Number903-111
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerRADIOMETER MEDICAL APS
Manufacturer AddressAKANDEVEJ 21 BRONSHOJ DK-2700 DA DK-2700


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2014-11-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.