RTM 600 TREADMILL 950-280

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-11-15 for RTM 600 TREADMILL 950-280 manufactured by Biodex Medical Systems, Inc..

Event Text Entries

[5049903] Patient fell off treadmill and dislocated her shoulder.
Patient Sequence No: 1, Text Type: D, B5


[12538780] Biodex was notified (b)(4) 2014 about the problem. We were told the patient was instructed how to use the treadmill and then was not supervised. With pressing the speed button continuously the mill sped up too fast and patient fell off. They did not use the emergency stop button, stop button on the display or the safety lanyard. All of which will disable unit. Biodex is waiting to analyze the unit. Response to 4h-3: biodex medical systems is waiting to hear from the customer so we can go in and evaluate the system, determine the cause and make repairs, if necessary.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2431314-2014-00001
MDR Report Key4261723
Report Source05,06
Date Received2014-11-15
Date of Report2014-10-28
Date of Event2014-06-16
Date Mfgr Received2014-10-23
Device Manufacturer Date2008-02-01
Date Added to Maude2014-11-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA0
Event Location0
Manufacturer ContactCLYDE SCHLEIN, VP
Manufacturer Street20 RAMSAY RD.
Manufacturer CitySHIRLEY NY 119674704
Manufacturer CountryUS
Manufacturer Postal119674704
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRTM 600 TREADMILL
Generic NameTREADMILL
Product CodeIOL
Date Received2014-11-15
Model Number950-280
Catalog Number950-280
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBIODEX MEDICAL SYSTEMS, INC.
Manufacturer Address20 RAMSAY RD. SHIRLEY NY 11967 US 11967


Patients

Patient NumberTreatmentOutcomeDate
10 2014-11-15

© 2026 FDA.report