SEAL RITE MASK WITH ONE WAY VALVE 120

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00 report with the FDA on 2014-11-19 for SEAL RITE MASK WITH ONE WAY VALVE 120 manufactured by Enter Medical Corp.

Event Text Entries

[21446575] The issues with the suspect medical device, seal rite masks, were initially reported to the importer of this device, the (b)(4), on (b)(6) 2014 by a customer who purchased and used this mask in a (b)(4) training course. Below are the initial reporter's own words: "i purchased the seal rite masks to replace my existing seal easy masks. I used them this past weekend in (b)(4) training and was very disappointed. Immediately the masks began to come apart, would not stay inflated, the valve to adjust air came apart on many masks, water seeped into the mask and some just leaked air constantly. " (b)(4) immediately stopped selling these masks and informed their contract mfr ((b)(4)) of the issues on 05/24/2014. The (b)(4) did their best to recall the defective medical devices sold. The (b)(4) and (b)(4) worked closely to rectify these defects. Since early june, 2014, all the defects have been rectified. Please reference mw5036511 and (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[21687927] The seal rite mask with one-way valve consists of a circular cushioned mask which can be easily inflated by hand and a one-way non-rebreathing valve. The intended use of the device is for the (b)(4) to provide ventilation to the pt. The design spec is provided by the (b)(4) and mfg by the contract mfr (b)(4). The first batch of the product was shipped in (b)(4) 2014 to the (b)(4), then sold to the customers. On (b)(4)2014, the (b)(4) received complaints from a customer. The (b)(4) passed this info to (b)(4) on 05/24/2014, stopped selling and recalled the defective masks immediately. After thorough analysis, the following three issues were discovered: the plastic connector piece on the bladder falls apart when impacted by a great force (jumping into water). The upper plastic piece in the inflation valve also comes off under great force. A red port cap on the oxygen attachment port of the one-way valve also comes off under great force. (b)(4) rectified the issues with following three fixes: put three points of pvc glue on the plastic connector to hold it in place so that it doesn't come off. Do high-frequency binding between upper and lower plastic pieces in the inflation valve so that they stay attached even under great force. Modify the mold of the port cap to decrease the circumference of the base circle and increase the thickness on the bottom so they do not slide off easily. Since 06/10/2014, all known issues have been rectified and new masks and valves have been shipped. The (b)(4) has done their best to recall the defective masks and replace them with new updated masks.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3006157842-2014-00001
MDR Report Key4262163
Report Source00
Date Received2014-11-19
Date of Report2014-05-24
Date of Event2014-05-23
Date Facility Aware2014-05-23
Report Date2014-05-24
Date Mfgr Received2014-05-24
Device Manufacturer Date2014-04-01
Date Added to Maude2014-11-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJOHN LEE
Manufacturer Street16F1 NO.16 JIAN BA ROAD CHUNG HO DISTRICT,
Manufacturer CityNEW TAIPEI CITY 23511
Manufacturer CountryTW
Manufacturer Postal23511
Manufacturer Phone282265221
Manufacturer G1ENTER MEDICAL CORP
Single Use3
Remedial ActionMA
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSEAL RITE MASK WITH ONE WAY VALVE
Product CodeBSJ
Date Received2014-11-19
Model Number120
Lot Number14015
OperatorOTHER
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerENTER MEDICAL CORP
Manufacturer AddressCHIA YI COUNTY TW


Patients

Patient NumberTreatmentOutcomeDate
10 2014-11-19

© 2026 FDA.report