AXIOM MULTIX MT 8395415 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,07 report with the FDA on 2014-11-13 for AXIOM MULTIX MT 8395415 NA manufactured by Siemens Medical Solutions Usa, Inc..

Event Text Entries

[5029257] A customer reported the 3d tube was stiff to move on the multix mt system and requested service. A siemens engineer was dispatched to the facility. Upon inspection the engineer noticed that het transverse roller was hanging over the edge of the rail. The securing pins for the transverse bridge assembly came out and were not locking the transverse rail to the carriage unstable. Although there have been no report of any injuries in this event, the whole 3d assembly could potentially crash down during an examination. The reported event occurred in the (b)(6).
Patient Sequence No: 1, Text Type: D, B5


[12574136] The concerned system was repaired by the engineer and returned to service. A customer site visit was planned by the development expert. The reported issue is under investigation and a supplemental report will be submitted once add'l info has been received. Customer's address: (b)(6). This report was filed (b)(6) 2014.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2240869-2014-08382
MDR Report Key4269192
Report Source01,07
Date Received2014-11-13
Date of Report2014-10-21
Date of Event2014-10-14
Date Mfgr Received2014-10-21
Date Added to Maude2014-11-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactANASTASIA MASON
Manufacturer Street51 VALLEY STREAM PKWY. MS D-02
Manufacturer CityMALVERN PA 193551406
Manufacturer CountryUS
Manufacturer Postal193551406
Manufacturer Phone6102194834
Manufacturer G1SIEMENS AG
Manufacturer StreetSIEMENSSTRASSE 1
Manufacturer CityFORCHHEIM 91301
Manufacturer CountryGM
Manufacturer Postal Code91301
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAXIOM MULTIX MT
Generic NameRADIOLOGIC TABLE
Product CodeIZZ
Date Received2014-11-13
Model Number8395415
Catalog NumberNA
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS MEDICAL SOLUTIONS USA, INC.
Manufacturer Address51 VALLEY STREAM PKWY. MS D-02 MALVERN PA 19355140 US 19355 1406


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2014-11-13

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.