CAPILLARY CAPS 10311054

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06,07 report with the FDA on 2014-11-24 for CAPILLARY CAPS 10311054 manufactured by Siemens Healthcare Diagnostics.

Event Text Entries

[5030884] Customer reported that capillary caps were not fitting well to glass capillaries and causing glass capillaries to break. Customer also reported that operator had minor cut on the finger due to this incident. There was no report of medical intervention due to this incident.
Patient Sequence No: 1, Text Type: D, B5


[12569893] Customer indicated that they do not have any capillaries left from a lot that had not been breaking for investigation. This issue is under investigation. The cause for the event is unknown.
Patient Sequence No: 1, Text Type: N, H10


[34037869] (b)(4). Additional information:siemens investigation identified that a material change made by the supplier has resulted in capillary caps that are more difficult to apply to the capillaries. An urgent field safety notice #(b)(4) was sent by e-mail to all affected siemens healthcare diagnostics regional offices on january 22, 2015 both in the united states and elsewhere for communication with affected customers. Please note: these capillary caps are an "off the shelf" item. Siemens purchases them from a distributor and repackages them.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1217157-2014-00178
MDR Report Key4274206
Report Source06,07
Date Received2014-11-24
Date of Report2014-10-29
Date of Event2014-10-17
Date Mfgr Received2015-01-23
Date Added to Maude2014-12-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactSTEVEN ANDBERG
Manufacturer StreetNORTHERN ROAD CHILTON INDUSTRIAL ESTATE
Manufacturer CitySUDBURY MA 02062CO10
Manufacturer CountryUK
Manufacturer Postal02062CO10
Manufacturer Phone7812693655
Manufacturer G1VITREX MEDICAL A/S
Manufacturer StreetVASEKAER 6-8 2730 HERLEV.
Manufacturer CountryDA
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction Number1217157-01-22-2015-001-C
Event Type3
Type of Report3

Device Details

Brand NameCAPILLARY CAPS
Generic NameCAPILLARY CAPS
Product CodeGIO
Date Received2014-11-24
Catalog Number10311054
Lot Number1710264N
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer Address511 BENEDICT AVENUE TARRYTOWN 10591 US 10591


Patients

Patient NumberTreatmentOutcomeDate
10 2014-11-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.