[5034928] Tmm notified via voice mail on (b)(6) 2014 by (b)(4) that an unattended icu pt fell out the front (foot section) of the chair. The root cause was the pt egressed from the front the device instead of a side ingress/egress, which is standard protocol. The chair tipped up and the pt fell on the floor. Exact circumstances of how this occurred are unk as no medical personnel witness the adverse event. Advocate stated that a subsequent medical eval yielded no serious injuries.
Patient Sequence No: 1, Text Type: D, B5


[12585970] (b)(4) stated a preliminary investigation by advocate biomed was completed and felt the chair was working as it should. A tmm representative promptly reviewed the equipment on 11/05/2014 and a confirmatory inspection by also found it to be in good working condition. As this pt was unattended, it is not clearly understand of how the pt was using the device. Tmm clearly states in all of our training materials that the ingress/egress protocol is from the side, and not the front. Third party, (b)(4) load testing verifies the safe and effective stability testing of the device.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004082462-2014-00003
MDR Report Key4279073
Report Source06
Date Received2014-11-20
Date of Report2014-11-14
Date of Event2014-10-23
Date Mfgr Received2014-11-03
Device Manufacturer Date2005-12-01
Date Added to Maude2014-11-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street1441 WOLF CREEK TRAIL
Manufacturer CitySHARON CENTER OH 44274
Manufacturer CountryUS
Manufacturer Postal44274
Manufacturer Phone3302394192
Single Use0
Remedial ActionIN
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRANSMOTION MEDICAL, INC.
Generic NameTMM4 MULTI-PURPOSE STRETCHER-CHARI
Product CodeGBB
Date Received2014-11-20
Model NumberTMM4BT
Catalog NumberTMM4BT
Lot Number0757
OperatorLAY USER/PATIENT
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTRANSMOTION MEDICAL INC.
Manufacturer Address1441 WOLF CREEK TRAIL SHARON CENTER OH 44274 US 44274


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2014-11-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.