GAMBRO WATER PURIFICATION SYSTEM 100474

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2014-11-26 for GAMBRO WATER PURIFICATION SYSTEM 100474 manufactured by Mar Cor Purification.

Event Text Entries

[15313763] Mar cor purification installed a 10 cubic foot carbon tank in the clinic's pretreatment portion of the reverse osmosis water system in (b)(6) of 2013. Mcp was notified on (b)(6) 2014 that the carbon tank feed side union leaked and sprayed water on the ceiling. The water was deflected onto the top of the clinics cwp reverse osmosis water system causing the electronics and main control panel to become wet. This caused the cwp to stop generating reverse osmosis water resulting in the delay of patient treatment.
Patient Sequence No: 1, Text Type: D, B5


[15645561] Due to the water saturation of the electronics and main cwp ro control panel the customer was advised to replace the entire cwp ro control panel. Patients were not able to dialyze until deionized water tanks were brought online, after which all patients were able to be dialyzed. There were no reports of patient illness or injury as a result of this incident. Mar cor purification will continue to monitor this complaint within the complaint system.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3019131-2014-00032
MDR Report Key4282249
Report Source06
Date Received2014-11-26
Date of Report2014-11-26
Date of Event2014-10-27
Date Mfgr Received2014-10-27
Date Added to Maude2014-12-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. MEGAN DICKEY
Manufacturer Street14550 28TH AVENUE NORTH
Manufacturer CityMINNEAPOLIS MN 55428
Manufacturer CountryUS
Manufacturer Postal55428
Manufacturer Phone7632103866
Manufacturer G1MAR COR PURIFICATION
Manufacturer Street14550 28TH AVENUE NORTH
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal Code55447
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGAMBRO WATER PURIFICATION SYSTEM
Generic NameWATER PURIFICATION SYSTEM FOR HEMODIALYSIS
Product CodeFIP
Date Received2014-11-26
Model Number100474
ID Number1000136
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMAR COR PURIFICATION
Manufacturer Address14550 28TH AVENUE NORTH MINNEAPOLIS MN 55428 US 55428


Patients

Patient NumberTreatmentOutcomeDate
10 2014-11-26

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.