[16179900] Patient underwent cataract removal with toric iol implantation on (b)(6) 2014. First day post op eye exam was normal. On one week post-operative examination with use of the slit lamp and with the eye dilated, a small cylinder shaped piece of metal was identified near the lens. The patient's eye exam was otherwise normal and she denied visual difficulties. The patient underwent outpatient surgery on (b)(6) 2014 to remove the metal fragment without difficulty. After removal, examination of the metal fragment under microscope, showed that it was part of the tip for the mandelburg irrigating rotator which is used when implanting a toric iol and was used on this patient a week ago. Key points that link the metal piece to the device are the following: the mandelburg irrigating rotator when tested on (b)(6) 2014 did not function as per usual. It only irrigated out of the end of the device rather than through the two side ports. The size/diameter of the metal object exactly matched the size/diameter of the end of the mandelburg irrigating rotator. The metal object was the same blue color as the mandelburg irrigating rotator tip.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5039305
MDR Report Key4284327
Date Received2014-11-24
Date of Report2014-11-12
Date of Event2014-01-11
Date Added to Maude2014-12-01
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameMST
Generic NameMANDELBURG IRRIGATING ROTATOR
Product CodeHNI
Date Received2014-11-24
Model NumberNOT APPLICABLE
Catalog NumberUNK
Lot NumberNOT APPLICABLE
ID NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Sequence No1
Device Event Key0
ManufacturerMICROSURGICAL TECHNOLOGY, INC.
Manufacturer AddressREDMOND WA US


Patients

Patient NumberTreatmentOutcomeDate
10 2014-11-24