TRANSMOTION MEDICAL, INC. TMM5-TB

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2014-11-20 for TRANSMOTION MEDICAL, INC. TMM5-TB manufactured by Transmotion Medical, Inc..

Event Text Entries

[5227156] There was no pt event therefore, we are leaving the date above as "n/a". A service call was received on 09/18/2014 to address the product moving on its own. It was decided on 10/21/2014, that we should launch an mdr as good measure for reporting purposes. This mdr is being filed to notify of the potential for injury due to the chair moving on its own due to a design issue with the hand control pendant. Again, there was no pt nor a pt incident involved.
Patient Sequence No: 1, Text Type: D, B5


[12584187] The immediate action taken was to replace all older styler pendants with newly designed pendants for all devices in the facility. There was not an adverse event. No pt was on the device. Transmotion medical inc. Has designed a new pendant which was released in may of 2013. The new pendant is considerably more durable than the older style pendants. The new pendant has been in production for 17 months and there have been no reports of the new pendant causing the device to move on its own. Additionally, in november 2014, transmotion medical, inc launched a product recall due to previous customer complaint and corresponding mdr data.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004082462-2014-00002
MDR Report Key4286539
Report Source06
Date Received2014-11-20
Date of Report2014-11-14
Device Manufacturer Date2007-05-01
Date Added to Maude2014-12-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street1441 WOLF CREEK TRAIL
Manufacturer CitySHARON CENTER OH 44274
Manufacturer CountryUS
Manufacturer Postal44274
Manufacturer Phone3302394192
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRANSMOTION MEDICAL, INC.
Generic NameTMM5 MOBILE SURGICAL STRETCHER-CHAIR
Product CodeGBB
Date Received2014-11-20
Model NumberTMM5-TB
Catalog NumberTMM5-TB
Lot Number1338, 1355-1362
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTRANSMOTION MEDICAL, INC.
Manufacturer Address1441 WOLF CREEK TRAIL SHARON CENTER OH 44274 US 44274


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2014-11-20

© 2026 FDA.report