ACE HALO SYSTEM 1374

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2002-11-26 for ACE HALO SYSTEM 1374 manufactured by Depuy Acromed, Inc..

Event Text Entries

[251396] It was reported to depuy acromed that a patient wearing a universal tong during a mri scan, complained about pain and burning around their head. The scan was stopped and there were no adverse consequences to the patient. Further information revealed that the hospital recently upgraded the mri computer system which caused it to be more powerful than what the tong was designed to withstand. A complaint history analysis was performed and no other complaints have been reported for this part number. The part was not returned for evaluation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1526439-2002-00074
MDR Report Key430048
Report Source07
Date Received2002-11-26
Date of Report2002-11-26
Date Mfgr Received2002-11-14
Date Added to Maude2002-12-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street325 PARAMOUNT DRIVE
Manufacturer CityRAYNHAM MA 02767
Manufacturer CountryUS
Manufacturer Postal02767
Manufacturer Phone5088808100
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameACE HALO SYSTEM
Generic NameUNIVERSAL TONG
Product CodeHAX
Date Received2002-11-26
Model NumberNA
Catalog Number1374
Lot NumberUNK
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeNA
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key419037
ManufacturerDEPUY ACROMED, INC.
Manufacturer Address325 PARAMOUNT DRIVE RAYNHAM MA 02767 US
Baseline Brand NameACE UNIVERSAL TONG KIT
Baseline Generic NameHALO, TONGS, PINS
Baseline Catalog No1374
Baseline Device FamilyACE HALO SYSTEMS
Baseline Shelf Life [Months]NA
Baseline PMA FlagN
Baseline 510K PMNY
Premarket NotificationK954069
Baseline PreamendmentN
Baseline TransitionalN
510k ExemptN


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2002-11-26