HELIOS 3000 WALL MOUNTED LIGHT HL3WL

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2014-11-25 for HELIOS 3000 WALL MOUNTED LIGHT HL3WL manufactured by Pelton & Crane.

Event Text Entries

[5211191] It was reported that a dental professional was moving a pelton and crane helios 3000 dental light away after use when the arm assembly broke causing the light to fall down towards the ground. The light grazed the patient on the side of their head; however, the patient was not injured.
Patient Sequence No: 1, Text Type: D, B5


[12711621] Upon evaluation of the light arm rear knuckle assembly, it was observed that a portion of the knuckle had broken off the main knuckle assembly. The fracture of the knuckle started where the rotational stop pin is located. The rotational stop pin is designed to prevent the operator from rotating the light more than 360 degrees during use. It was also observed that the rotational stop pin was missing. The round counter bore hole where the stop pin is located was deformed in an oblong shape. This is indicative of force being applied to the rotational stop pin as the direction of the deformation was towards the section of the knuckle casting that had broken off. As a result, the rotational stop pin was exposed to an excessive force during use causing the failure of the knuckle assembly.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1017522-2014-00015
MDR Report Key4307650
Report Source08
Date Received2014-11-25
Date of Report2014-10-30
Date of Event2014-10-30
Date Mfgr Received2014-10-30
Device Manufacturer Date2010-05-01
Date Added to Maude2014-12-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactFRANK RAY, MANAGER
Manufacturer Street11727 FRUEHAUF DR.
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer Phone7045877227
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHELIOS 3000 WALL MOUNTED LIGHT
Generic NameDENTAL OPERATING LIGHT
Product CodeEAZ
Date Received2014-11-25
Model NumberHL3WL
Catalog NumberHL3WL
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPELTON & CRANE
Manufacturer Address11727 FRUEHAUF DR. CHARLOTTE NC 28273 US 28273


Patients

Patient NumberTreatmentOutcomeDate
10 2014-11-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.