SMART MONITOR 2 PS 1014557

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2014-08-26 for SMART MONITOR 2 PS 1014557 manufactured by Philips Respironics -chmv.

Event Text Entries

[5212097] Philips healthcare received a complaint of a battery draining too quickly. The complaint issue alleged by the customer was not observed during the repair evaluation of the device at the manufacturer's service ctr, however, a finding of "faulty alarm module" not related to the customer's complaint was observed. The secondary alarm did alarm as expected but we are reporting the on the faulty alarm module preventatively. The module will be sent to the supplier and a f/u report will be filed once the investigation is complete.
Patient Sequence No: 1, Text Type: D, B5


[12639383] (b)(4) a f/u report will be submitted once the investigation is complete. Pma 510(k): k032403, k061256.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1218950-2014-07405
MDR Report Key4330095
Report Source06
Date Received2014-08-26
Date of Report2014-08-01
Date Mfgr Received2014-08-01
Date Added to Maude2014-12-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNANCY ATAIDE
Manufacturer Street3000 MINUTEMAN RD.
Manufacturer CityANDOVER MA 01810
Manufacturer CountryUS
Manufacturer Postal01810
Manufacturer Phone9786597429
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSMART MONITOR 2 PS
Generic NameNPF
Product CodeNPF
Date Received2014-08-26
Model Number1014557
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerPHILIPS RESPIRONICS -CHMV
Manufacturer Address191 WYNGATE DIVE MONROEVILLE PA 15146 US 15146


Patients

Patient NumberTreatmentOutcomeDate
10 2014-08-26