MICROSELECTRON MHDR DIGITAL

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,05,06 report with the FDA on 2014-12-16 for MICROSELECTRON MHDR DIGITAL manufactured by Nucletron.

Event Text Entries

[15321846] The patient visited the breast surgeon for skin redness, two weeks after a savi applicator breast treatment. The surgeon reviewed the treatment plan and determined that the catheter was reconstructed backwards with the dose heavily weighted toward the connector end of the savi applicator. This results in a dose given at certain proximity to the skin which can cause skin redness. Nucletron did not receive data to determine the severity of the event. The savi applicator was connected to a nucletron microselectron afterloader and the treatment planning was performed with nucletron oncentra brachytherapy treatment planning software.
Patient Sequence No: 1, Text Type: D, B5


[15552861] The user was consulted on how such a case can be avoided. The user was advised this can be prevented with features available for this in oncentra brachy. No specific details were provided by the user to perform further analysis. No information received from hospital.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9611894-2014-00004
MDR Report Key4332511
Report Source00,05,06
Date Received2014-12-16
Date of Report2014-12-16
Date of Event2014-11-21
Date Mfgr Received2014-11-21
Date Added to Maude2014-12-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer StreetLINAC HOUSE FLEMING WAY
Manufacturer CityCRAWLEY, WEST SUSSEX RH109RR
Manufacturer CountryUK
Manufacturer PostalRH10 9RR
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMICROSELECTRON MHDR
Generic NameSYSTEM, APPLICATOR, RADIONUCLIDE, REMOTE-CONTROLLED
Product CodeJAQ
Date Received2014-12-16
Model NumberDIGITAL
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerNUCLETRON
Manufacturer AddressWAARDGELDER 1 VEENENDAAL, NL


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2014-12-16