MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2002-12-13 for FORMASPLINT PB 105 * manufactured by Thermo-electric Co..

Event Text Entries

[299819] Drain heater coil overheated, resulting in production of smoking within unit. Visual check shows evidence of fire or burns. Operator notified fire department at time of incident. Fire department instructed to bring for inspection. Design flaw evident, as conditions has high likelihood of recurrence even after repair. Unit assembled with fiberglass insulation in contact with drain heater assembly and then covered. Suggest manufacturer re-design drain heater to isolate heating element from any flammable materials. Also, a timer should be placed in circuit to prevent power from being applied for an extended time to drain heater. Strongly recommend manufacturer design circuit breaker or fuse into unit to interrupt power in case of malfunction.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW1027003
MDR Report Key433601
Date Received2002-12-13
Date of Report2002-12-13
Date Added to Maude2002-12-23
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA0
Report to FDA0
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameFORMASPLINT
Generic NameBATH, PARAFFIN
Product CodeIMC
Date Received2002-12-13
Model NumberPB 105
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityN
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key422544
ManufacturerTHERMO-ELECTRIC CO.
Manufacturer Address455 ROUTE 30 IMPERIAL PA 15126 US