WAVESCAN FOURIER ALGORITHM 0070-1006

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-12-17 for WAVESCAN FOURIER ALGORITHM 0070-1006 manufactured by Abbott Medical Optics.

Event Text Entries

[5176645] Account reported the computer was not functioning and the power supply fans were not operating. Customer saw smoke coming from back of the computer upon powering up the wavescan for use. System shut down.
Patient Sequence No: 1, Text Type: D, B5


[12702253] Field service specialist visited the account and smelled odor from enclosed power supply on computer. Replaced power supply assembly and tested the equipment. System meets amo specifications. All pertinent information available to abbott medical optics has been submitted. Placeholder.
Patient Sequence No: 1, Text Type: N, H10


[23244448] Corrections: health professional is being corrected to yes for the initial medical device report (mdr). Health professional was not included in the initial mdr. Evaluation: operation problem for results is being corrected to power supply for the initial mdr. All pertinent information available to the abbott medical optics has been submitted.
Patient Sequence No: 1, Text Type: N, H10


[23244809] All pertinent information available to abbott medical optics has been submitted.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3006695864-2014-00550
MDR Report Key4338687
Report Source05,06
Date Received2014-12-17
Date of Report2014-11-26
Date of Event2014-11-26
Date Mfgr Received2015-04-20
Device Manufacturer Date2006-09-18
Date Added to Maude2014-12-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMRS. VALERIE SEDZICKI
Manufacturer Street1700 EAST ST. ANDREW PLACE SA3M - C3051
Manufacturer CitySANTA ANA CA 92705
Manufacturer CountryUS
Manufacturer Postal92705
Manufacturer Phone7142478567
Manufacturer G1ABBOTT MEDICAL OPTICS INC.
Manufacturer Street510 COTTONWOOD DRIVE
Manufacturer CityMILPITAS CA 95035
Manufacturer CountryUS
Manufacturer Postal Code95035
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWAVESCAN FOURIER ALGORITHM
Generic NameREFRACTIVE MEASUREMENT
Product CodeHKO
Date Received2014-12-17
Model Number0070-1006
Catalog Number0070-1006
OperatorPHYSICIAN
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MEDICAL OPTICS
Manufacturer AddressSANTA ANA CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2014-12-17