BRACHY STAR SEED IMPLANT NEEDLE 818205

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2014-12-12 for BRACHY STAR SEED IMPLANT NEEDLE 818205 manufactured by Bard Reynosa S.a. De C.v..

Event Text Entries

[5201383] It was reported that the needle broke while in the pt. The needle contained 6 seeds when it broke during the implant. The needle was in the pt when it broke but they were able to retrieve both parts of the needle and the seeds were still in the needle. Per additional info received the needle was being manipulated between the pt and grid with a diddler" (tool used to help guide the needle tip to the target area of the prostate). The needle was in the pt's perinium approximately 1-2 cm when the breakage occurred leaving 2-3 cm protruding outside of the pt's body. The physician was able to grasp the needle with forceps and pull out the broken portion from the pt. The doctor performed a fluoroscopy to ensure that no fragments remained in the pt. No fragments were noted.
Patient Sequence No: 1, Text Type: D, B5


[12701537] The investigation is still in progress. Once the investigation is complete a supplemental report will be filed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1018233-2014-00352
MDR Report Key4349338
Report Source07
Date Received2014-12-12
Date of Report2014-11-14
Date Mfgr Received2014-11-14
Date Added to Maude2014-12-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCHRISTY LEWIS
Manufacturer Street8195 INDUSTRIAL BLVD.
Manufacturer CityCOVINGTON GA 30014
Manufacturer CountryUS
Manufacturer Postal30014
Manufacturer Phone7707846100
Manufacturer G1BARD REYNOSA S.A. DE C.V.
Manufacturer StreetBLVD MONTEBELLO #1 PARQUE INDUSTRIAL COLONIAL
Manufacturer CityREYNOSA, TAMAULIPAS
Manufacturer CountryMX
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameBRACHY STAR SEED IMPLANT NEEDLE
Product CodeIWJ
Date Received2014-12-12
Model NumberNA
Catalog Number818205
Lot NumberREYF1142
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBARD REYNOSA S.A. DE C.V.
Manufacturer AddressREYNOSA, TAMAULIPAS MX


Patients

Patient NumberTreatmentOutcomeDate
10 2014-12-12

© 2026 FDA.report