SPIRIT 1700 DENTAL CHAIR SP17 E1704

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2014-12-19 for SPIRIT 1700 DENTAL CHAIR SP17 E1704 manufactured by Pelton & Crane.

Event Text Entries

[5237590] A dental professional was reclining a patient for treatment in a pelton & crane sp17 dental chair when the lift assembly broke resulting in the seat assembly to fall backwards towards the floor. The patient did not fall out of the chair. There were no injuries reported.
Patient Sequence No: 1, Text Type: D, B5


[12678952] Upon evaluation it was identified that the two welded support block on the front truss assembly had failed as a result of improper welding from our supplier as it was not welded to manufacturers specifications. Pelton & crane has already contacted the supplier of the front truss assembly and informed then of this issue. The supplier informed pelton & crane that they will update fqc document to include a conformity check of weldments to ensure front truss assemblies are manufactured to pelton & crane specifications.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1017522-2014-00016
MDR Report Key4360995
Report Source08
Date Received2014-12-19
Date of Report2014-12-04
Date of Event2014-12-03
Date Mfgr Received2014-12-04
Device Manufacturer Date2012-04-01
Date Added to Maude2015-01-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactFRANK RAY
Manufacturer Street11727 FRUEHAUF DR.
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer Phone7045877227
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSPIRIT 1700 DENTAL CHAIR
Generic NameDENTAL CHAIR
Product CodeKLC
Date Received2014-12-19
Returned To Mfg2014-12-15
Model NumberSP17
Catalog NumberE1704
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPELTON & CRANE
Manufacturer Address11727 FRUEHAUF DR. CHARLOTTE NC 28273 US 28273


Patients

Patient NumberTreatmentOutcomeDate
10 2014-12-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.