ABG-HM-1; HUMMI MICRO DRAW BLOOD TRANSFER DEVI

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,08 report with the FDA on 2014-08-04 for ABG-HM-1; HUMMI MICRO DRAW BLOOD TRANSFER DEVI manufactured by Hummingbird Med.

Event Text Entries

[5119447] It was reported that a baxter t-connector had leaked, after being used to access arterial blood from a baby with the abg-hm-1 (hummi). The nurse stated that the baby lost 8ml of blood and required a blood transfusion due to the blood loss, since it had occurred on a low birth weight baby. The nurse mentioned that the septum had the appearance of being somewhat "pushed in. "
Patient Sequence No: 1, Text Type: D, B5


[12630668] Hummingbird med simulation test indicates user error (not following ifu and in-service training) by inserting the hummi device inappropriately into the perimeter of the septum. Improper insertion of any blunt cannula may deform the septum and allow fluids to leak out between the septum and sidewall of the device. Direct communication with the reporting nurse at the hospital indicated the hospital has stopped using hummi product until an appropriate t-connector from another supplier is available. Contact nurse could not supply further info. The failure of the t-connector was reported by the nurse to baxter. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3009382876-2014-00001
MDR Report Key4366749
Report Source05,08
Date Received2014-08-04
Date of Report2014-08-04
Date of Event2014-07-22
Date Mfgr Received2014-07-25
Date Added to Maude2014-12-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJORGE HAIDER
Manufacturer Street2797 PRADO LN
Manufacturer CityMARIETTA GA 30066
Manufacturer CountryUS
Manufacturer Postal30066
Manufacturer Phone7605856525
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameABG-HM-1; HUMMI MICRO DRAW BLOOD TRANSFER DEVI
Generic NameSYSTEM, BLOOD COLLECTION, VACUUM-ASS
Product CodeKST
Date Received2014-08-04
Model NumberABG-HM-1
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerHUMMINGBIRD MED
Manufacturer Address2797 PRADO LN MARIETTA GA 30066 US 30066


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2014-08-04

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.