PORT ACCESS, KIT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,06,07 report with the FDA on 2014-09-03 for PORT ACCESS, KIT manufactured by Medical Action Industries, Inc..

Event Text Entries

[22067351] Please refer to importer report#: (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[22269834] On (b)(4) 2014, medical action industries (mai) received a complaint via a sales representative that a (b)(4) blunt cannula non-coring safety needle was bent as a nurse was attempting to access the portacath. Mai does not mfr the defective needle, but does mfr the port access kit, 59043c (lots # 187966, 185508 and 189211) that contains the needle. Smiths medical mfrs the needle. Immediately upon receipt of the complaint, mai opened complaint numbers # 200016236, 200016237 and 200016238, one for each complaint lot, and began the investigation. On 08/04/2014, mai received medwatch # (b)(4) from fda. This specific needle lot # 34x136 was only used in the kit complaint lots. Mai's inventory of the specific needle lot involved and complaint kit lots was exhausted upon receipt of this complaint. Review of complaint history shows we have no additional complaints on the complaint lots. In addition, this needle is used in other kits with no complaints in regard to this concern. We do not see a trend in regard to this issue and determined the complaint to be an isolated incident. Our supplier, smiths medical, was notified of this issue, and was given samples from the customer to further address the issue. Smiths medical responded directly to the customer on (b)(4) 2014 that the samples received were within specifications of their critical diameter size dimension of the trocar/cannulas, and the bending observed was consistent with the gripper micro having been subjected to a lateral force rather than straight down. Per their instructions for use, the force should be applied at 90 degrees. Although our supplier confirmed the complaint, they found no evidence to suggest the event was caused from an intrinsic defect in the product or a mfg issue. Smiths medical's supplier eval summary is also attached. To date, all kit complaint lots have been exhausted from all facilities involved except for 15 cases of lot #189211 which has been quarantined at mai's distributor. Smiths medical has just notified the end user that they are confident there is no risk in putting the complaint needle and kits back into distribution. Mai has entered the complaint into our database and will continue to track for any related trends.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1030451-2014-00008
MDR Report Key4369847
Report Source00,06,07
Date Received2014-09-03
Date Mfgr Received2014-08-04
Date Added to Maude2015-01-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location0
Manufacturer ContactDEBORAH KLOOS
Manufacturer Street25 HEYWOOD RD.
Manufacturer CityARDEN NC 28704
Manufacturer CountryUS
Manufacturer Postal28704
Manufacturer Phone8286818820
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Generic NamePORT ACCESS, KIT
Product CodeLRS
Date Received2014-09-03
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDICAL ACTION INDUSTRIES, INC.
Manufacturer Address25 HEYWOOD RD. ARDEN NC 28704 US 28704


Patients

Patient NumberTreatmentOutcomeDate
10 2014-09-03