ON (B)(6) 2014, (B)(6) REPORTED THAT THE MASTER STATIONS OF THE NURSE CALL SYSTEM WERE ANNUNCIATING CALLS FROM SOME STATIONS, AND THIS DID NOT GENERATE A 'FAULT' CONDITION. AN UPDATED VERSION OF SOFTWARE WAS INSTALLED ON THE SYSTEM (B)(6) 2014. TO DATE THE PROBLEM HAS NOT RE-OCCURRED.
N
Patient 1
ROOT CAUSE: WE BELIEVE A COUNTING SEMAPHORE IN THE TC400 CODE CAUSED THE CENTRAL EQUIPMENT TO STOP POLLING. WE DETERMINED THIS FROM THE DATA RECEIVED FROM OUR DIAGNOSTICS TOOLS ON-SITE. CORRECTIVE ACTION: WE DETERMINED THROUGH TESTING THAT THE COUNTING SEMAPHORE IS NO LONGER NEEDED AND REMOVED IT FROM THE SYSTEM FIRMWARE. VALIDATION: BECAUSE WE'VE NOT BEEN ABLE TO DUPLICATE THE PROBLEM IN HOUSE, WE PLACED THE CODE ON-SITE AT (B)(6). TO DATE THE ORIGINAL PROBLEM HAS NOT RE-OCCURRED. PREVENTIVE ACTION: WE ADDED 'POLLING FAULT' CODE TO THE OS SOFTWARE. THIS WILL ALERT THE USER SHOULD THE NURSE CALL SYSTEM AGAIN BECOME UNRESPONSIVE.