* 5892

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2014-12-10 for * 5892 manufactured by Mizuho Orthopedic Systems, Inc.

Event Text Entries

[5285748] In the or, the sedated patient was moved from a supine position on a gurney to a osi orhtopedic table in prone position. Approximately 30 seconds after being placed on the table the front left wheel fell off and one side of the table dropped to the ground (approximately 4 inches). This sudden drop caused the table to tilt while the patient was on it. Staff quickly grabbed the table and held it in place, protecting the patient, until a gurney was brought in to transfer the patient. There was no harm to the patient and another orthopedic table was brought in and surgery was successfully completed. Biomed inspection of the broken wheel revealed the stem in the caster (wheel) had sustained damage at some time causing the flange that securely holds the stem to the wheel to break off. Unfortunately this area is not an inspection point during pm. Prior to the wheel falling off the table was moving without any wobbling or difficulty. ======================manufacturer response for osi orthopedic table, (brand not provided) (per site reporter). ======================they will inspect the table.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number4371782
MDR Report Key4371782
Date Received2014-12-10
Date of Report2014-12-10
Date of Event2014-12-05
Report Date2014-12-10
Date Reported to FDA2014-12-10
Date Reported to Mfgr2014-12-30
Date Added to Maude2014-12-30
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand Name*
Generic NameTABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED
Product CodeJEB
Date Received2014-12-10
Model Number5892
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityY
Device Sequence No1
Device Event Key0
ManufacturerMIZUHO ORTHOPEDIC SYSTEMS, INC
Manufacturer Address30031 AHERN AVE UNION CITY CA 94587 US 94587


Patients

Patient NumberTreatmentOutcomeDate
10 2014-12-10

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.