HELIOS 3000 CABINET MOUNTED LIGHT HL3CB

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-12-23 for HELIOS 3000 CABINET MOUNTED LIGHT HL3CB manufactured by Pelton And Crane.

Event Text Entries

[5289183] A dental professional was pushing a pelton and crane helios 3000 light away from use when the light arm assembly came apart and fell down towards the ground. The light head hit the pt on their knee. There were no injuries reported.
Patient Sequence No: 1, Text Type: D, B5


[12653284] Upon visual inspection by the local pelton and crane distributor, it was determined that the locking key and locking key screw were not installed on the helios 3000 dental light arm. The function of the locking key and locking key screw is to secure the light arm assembly together. The device was not returned to pelton and crane but rather it was serviced and placed back in use at the doctor's office and has been tested and is now functioning to manufacturer's specifications. A review of pelton and crane's helios 3000 dental light installation manual clearly documents and illustrates how to properly install the locking key and locking key screw during installation of the dental light. Pelton and crane reviewed the installation issue with the distributor.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1017522-2014-00017
MDR Report Key4377406
Report Source05,06
Date Received2014-12-23
Date of Report2014-12-09
Date of Event2014-12-08
Date Mfgr Received2014-12-09
Device Manufacturer Date2011-01-01
Date Added to Maude2015-01-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactFRANK RAY
Manufacturer Street11727 FRUEHAUF DR.
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer Phone7045877227
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHELIOS 3000 CABINET MOUNTED LIGHT
Generic NameDENTAL LIGHT
Product CodeEAZ
Date Received2014-12-23
Model NumberHL3CB
Catalog NumberHL3CB
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPELTON AND CRANE
Manufacturer Address11727 FRUEHAUF DR. CHARLOTTE NC 28273 US 28273


Patients

Patient NumberTreatmentOutcomeDate
10 2014-12-23

© 2026 FDA.report