SIMPLEXA FLU A/B AND RSV DIRECT ASSAY MOL2650

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2014-12-29 for SIMPLEXA FLU A/B AND RSV DIRECT ASSAY MOL2650 manufactured by Focus Diagnostics, Inc..

Event Text Entries

[5118122] The customer reported that they are having too many invalid internal control values on the patient samples. No results were reported to the doctor as they were unable to obtain readings. Review of the customer's run database found four invalid samples in one run. The samples were valid upon repeat testing using the same lot of reagents. The product insert was revised to advise the end-user that "after each use store dad discs flat with the numbered foil side up. "
Patient Sequence No: 1, Text Type: D, B5


[12632913] Investigation determined that carry-over contamination from successive runs can occur if the end-user stores the dad (direct amplification disc) vertically (i. E. On edge). Storage of the dad in any vertical position after initial use has a potential of allowing fluid to travel back through the channels and contaminate other unused wedges. This may result in obtaining false positive results. The product insert warning and precautions section has been revised to include the following item: "after each use store dad discs flat with the numbered foil side up. " (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2023365-2014-00051
MDR Report Key4379428
Report Source05
Date Received2014-12-29
Date of Report2013-11-02
Date of Event2013-11-01
Date Mfgr Received2014-05-05
Device Manufacturer Date2013-09-01
Date Added to Maude2015-01-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactVALERIE CIMMARUSTI, CLS
Manufacturer Street11331 VALLEY VIEW ST.
Manufacturer CityCYPRESS CA 90630
Manufacturer CountryUS
Manufacturer Postal90630
Manufacturer Phone5622406484
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction NumberZ-1958-2014
Event Type3
Type of Report3

Device Details

Brand NameSIMPLEXA FLU A/B AND RSV DIRECT ASSAY
Generic NameIN-VITRO-DIAGNOSTIC TEST, INFLUENZA
Product CodeOCC
Date Received2014-12-29
Model NumberMOL2650
Catalog NumberMOL2650
Lot Number23784
ID NumberNA
Device Expiration Date2014-09-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerFOCUS DIAGNOSTICS, INC.
Manufacturer AddressCYPRESS CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2014-12-29

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.