YC-1800

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-12-23 for YC-1800 manufactured by Nidek Co., Ltd. Hamacho Plant.

Event Text Entries

[5240457] Nidek received a complaint from the customer on (b)(4) 2014. Customer reported that the laser yc-1800; sn (b)(4) has not been working properly. The laser energy is being delivered too diffusely. Customer mentioned that there is too much energy anterior and posterior to the center of focus.
Patient Sequence No: 1, Text Type: D, B5


[12633735] The device has not been returned to nidek. The inspection has not been done yet, so the results of evaluation are not available. Nidek contacted the customer to verify the complaint and also confirmed that at this point, none of the surgeons have reported a patient injury due to the laser. If additional significant information is received at a later date, a follow-up report will be submitted. Nidek considers this failure mode a reportable event as the device has malfunctioned and has a potential to cause or contribute to a serious injury if the malfunction were to recur.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002807715-2014-00045
MDR Report Key4379445
Report Source05,06
Date Received2014-12-23
Date of Report2014-12-03
Date of Event2014-01-01
Date Mfgr Received2014-12-03
Device Manufacturer Date2006-05-01
Date Added to Maude2015-01-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNEO YAMAGUCHI
Manufacturer Street47651 WESTINGHOUSE DRIVE
Manufacturer CityFREEMONT CA
Manufacturer CountryUS
Manufacturer Phone5103537785
Manufacturer G1NIDEK CO., LTD. HAMACHO PLANT
Manufacturer Street34-14, HIROISHI
Manufacturer CityGAMAGORI, AICHI 443-0038
Manufacturer CountryJA
Manufacturer Postal Code443-0038
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameYC-1800
Generic NameND-YAG LASER
Product CodeLXS
Date Received2014-12-23
Model NumberYC-1800
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerNIDEK CO., LTD. HAMACHO PLANT
Manufacturer Address34-14M HIROISHI GAMAGORI, AICHI 443-0038 JA 443-0038


Patients

Patient NumberTreatmentOutcomeDate
10 2014-12-23

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.