SIMPLEXA FLU A/B AND RSV DIRECT ASSAY MOL2650

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2014-12-29 for SIMPLEXA FLU A/B AND RSV DIRECT ASSAY MOL2650 manufactured by Focus Diagnostics, Inc..

Event Text Entries

[5237569] The customer initially reported that the had three false positive results. Subsequent information stated that there were 6 samples that tested positive and then came up negative. They sent the positive samples to their local public health laboratory and the public health lab found the samples to be negative. The state lab performed a molecular method (rt-pcr) for confirmation and stated that it is a modified fda test. The laboratory reported the false positive results to the doctor. This supplemental mdr is for the second of the six false positive results that were reported to the doctor. Five additional mdrs are being submitted remaining events.
Patient Sequence No: 1, Text Type: D, B5


[12612896] There were no more samples (patient specimen or reaction mix) that could be returned for testing. A retain sample from the reported lot (ln 24160) was tested, (b)(4). No false positive results were observed. One hundred percent of the ntc replicates were negative for all analytes and the rna internal control was detected. No problems were observed during testing. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2023365-2014-00040
MDR Report Key4379495
Report Source05
Date Received2014-12-29
Date of Report2014-02-18
Date of Event2014-02-18
Date Mfgr Received2014-02-18
Device Manufacturer Date2013-11-01
Date Added to Maude2015-01-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactVALERIE CIMMARUSTI, CLS
Manufacturer Street11331 VALLEY VIEW ST.
Manufacturer CityCYPRESS CA 90630
Manufacturer CountryUS
Manufacturer Postal90630
Manufacturer Phone5622406484
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameSIMPLEXA FLU A/B AND RSV DIRECT ASSAY
Generic NameIN-VITRO-DIAGNOSTIC TEST, INFLUENZA
Product CodeOCC
Date Received2014-12-29
Model NumberMOL2650
Catalog NumberMOL2650
Lot Number24160
ID NumberNA
Device Expiration Date2014-11-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerFOCUS DIAGNOSTICS, INC.
Manufacturer AddressCYPRESS CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2014-12-29

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.