XOMED? INSTRUMENTS 3722058

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,06,07 report with the FDA on 2015-01-06 for XOMED? INSTRUMENTS 3722058 manufactured by Xomed Microfrance Mfg.

Event Text Entries

[5369053] It was reported? Malfunction of locking system: broken tip. It is the first time to use after purchasing.? Follow-up communications obtained additional information. The tip break resulted in a fragment. There was no patient impact or injury. Another instrument was used to complete the procedure.
Patient Sequence No: 1, Text Type: D, B5


[12613183] This device is used for therapeutic purposes. (b)(4). The device has been received and pending product analysis. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[34340653] Product analysis was completed by the quality engineering team. As received condition: received 1 sample(s), part number 3722058. Equipment used: microscope (zeiss stemi 2000c between 0,65 to 5,0 magnification settings). Observations: when viewed under magnification, the tip was bent which would have resulted in the reported malfunction. Manufacturing and mishandling cannot be ruled out as potential causes. Based on the observations there are multiple potential causes and each is equally likely. There was no additional definitive evidence to indicate a root cause. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9680837-2015-00003
MDR Report Key4385305
Report Source01,06,07
Date Received2015-01-06
Date of Report2014-12-14
Date of Event2014-11-30
Date Mfgr Received2015-01-15
Device Manufacturer Date2012-07-01
Date Added to Maude2015-01-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCHARLOTTE AYALA
Manufacturer Street6743 SOUTHPOINT DR N
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9043328372
Manufacturer G1MEDTRONIC XOMED, INC.
Manufacturer Street6743 SOUTHPOINT DR N
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameXOMED? INSTRUMENTS
Generic NamePICK, MICROSURGICAL EAR
Product CodeJYT
Date Received2015-01-06
Returned To Mfg2014-12-18
Model Number3722058
Catalog Number3722058
Lot Number120704
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerXOMED MICROFRANCE MFG
Manufacturer AddressSAINT-AUBIN-LE-MONIAL BOURBON-L'ARCHAMBAULT 3160 FR 3160


Patients

Patient NumberTreatmentOutcomeDate
10 2015-01-06

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.