SURGICAL SUCKER/ SUMP SET 201135NS

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 99 report with the FDA on 2014-11-19 for SURGICAL SUCKER/ SUMP SET 201135NS manufactured by Novosci Corp..

Event Text Entries

[15904176] On (b)(6) 2014 an incident was reported to novosci by the distributor medtronic. From their reports, during an operation the tip of a tmp sucker came unattached during the use in a surgery. It was uncertain if the device was broken while in the patients chest. The product was partially saved and pictures were taken for investigation purposes. The perfusionist stated that there was no injury to patient due to the incident. A investigation will follow.
Patient Sequence No: 1, Text Type: D, B5


[16062852] Pictures of the device were sent. Only partial piece of sucker in pictures. Initial investigation shows that the bonding agent seemed to have worked properly. The failure point for the device was the tip/sucker connection. This is a natural flex point for this device. Based on the pictures the failure seems to be the result of too much force being applied to the tip of sucker causing a break. A direction for use is included for the sucker/sump sets. It directs that care should be used when applying force to the sucker. Any excessive force or improper use could cause the tip to break. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1625519-2014-00001
MDR Report Key4392400
Report Source99
Date Received2014-11-19
Date of Report2014-11-17
Date of Event2014-11-05
Date Mfgr Received2014-11-14
Device Manufacturer Date2013-12-01
Date Added to Maude2015-01-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street2021 AIRPORT RD
Manufacturer CityCONROE TX 77301
Manufacturer CountryUS
Manufacturer Postal77301
Manufacturer Phone2812103363
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSURGICAL SUCKER/ SUMP SET
Generic NameADULT HEX HANDLE GROOVED - TIP SUCKE
Product CodeDTS
Date Received2014-11-19
Model Number201135NS
Lot Number295044/294810
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerNOVOSCI CORP.
Manufacturer AddressCONROE TX US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2014-11-19