COLLIMATOR ACCESSORY LATCH EXL-8 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1996-10-21 for COLLIMATOR ACCESSORY LATCH EXL-8 * manufactured by Mitsubishi Intl Corp..

Event Text Entries

[19897867] Collimator accessory latch failed during gantry movement causing the wedge mount tray and block mount tray to fall to the floor. A pt was on the treatment table and the trays barely missed the pt. A svc bulletin was issued by mfr regarding a problem of this nature. It calls for the parts to be inspected at the time of a pm as a minimum. A pm was done seven days prior to this incident. This bulletin was issued 12/28/89 and there has been no attempt that rptr is aware of to correct the defect. (*)
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW1010122
MDR Report Key43973
Date Received1996-10-21
Date Added to Maude1996-10-24
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA0
Report to FDA0
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameCOLLIMATOR ACCESSORY LATCH
Generic NameCOLLIMATOR
Product CodeIYC
Date Received1996-10-21
Model NumberEXL-8
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key44910
ManufacturerMITSUBISHI INTL CORP.
Manufacturer AddressTWO TOWER BRIDGE 1 FAYETTE ST STE 100 CONSHOHOCKEN PA 19428 US


Patients

Patient NumberTreatmentOutcomeDate
10 1996-10-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.