[5309658] The system fault occurred during a procedure.
Patient Sequence No: 1, Text Type: D, B5


[12652356] The system fault occurred during a procedure. The system detected a potential hardware issue that would require a reboot to clear the message. The customer continued the case with a manual approach. No patient injury was reported and the system responded as designed when detecting a fault condition. Subsequent to the initial evaluation, additional review determined that in this specific case, the error message was incorrectly activated and no actual hardware fault existed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3006026430-2015-00001
MDR Report Key4398285
Report Source05,07
Date Received2015-01-06
Date of Report2015-01-06
Date of Event2014-03-18
Date Mfgr Received2014-03-18
Device Manufacturer Date2010-09-30
Date Added to Maude2015-01-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactRICK ALLEN, SR. DIRECTOR
Manufacturer Street800 E. MIDDLEFIELD RD.
Manufacturer CityMOUNTAIN VIEW CA 94043
Manufacturer CountryUS
Manufacturer Postal94043
Manufacturer Phone6504042731
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSENSEI X
Generic NameSENSEI X ROBOTIC CATHETER SYSTEM
Product CodeDXX
Date Received2015-01-06
Model Number08257
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHANSEN MEDICAL
Manufacturer AddressMOUNTAIN VIEW CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2015-01-06