HARMONY IQ 3600 INTEGRATION SYSTEM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2015-01-09 for HARMONY IQ 3600 INTEGRATION SYSTEM manufactured by Vts Medical Systems.

Event Text Entries

[5366593] The user facility reported during a patient procedure, one of their in-field monitors went out. No injuries or procedural delays or cancellations were reported.
Patient Sequence No: 1, Text Type: D, B5


[12744414] A steris field service technician arrived onsite, inspected the unit, and worked with a member of the ori team to test the facility wiring and integration system. The technician identified an arc within the unit. A short was occurring between the red power cable for the monitor and the grounding spring cable that is routed in the mesh cover as a part of the wiring bundle for the boom. When the cable came in contact with the spindle, the system experienced an electrical short. The technician replaced the cpu stack, the monitor cable, and the serial board. The two monitors will also be replaced to ensure the system will be operating according to specification.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1000404456-2015-00001
MDR Report Key4404300
Report Source06
Date Received2015-01-09
Date of Report2015-01-09
Date of Event2014-12-11
Date Mfgr Received2014-12-11
Date Added to Maude2015-01-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1VTS MEDICAL SYSTEMS
Manufacturer Street40 MELVILLE PARK RD
Manufacturer CityMELVILLE NY 11747
Manufacturer CountryUS
Manufacturer Postal Code11747
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHARMONY IQ 3600 INTEGRATION SYSTEM
Generic NameINTEGRATION SYSTEM
Product CodeLMD
Date Received2015-01-09
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerVTS MEDICAL SYSTEMS
Manufacturer Address40 MELVILLE PARK RD MELVILLE NY 11747 US 11747


Patients

Patient NumberTreatmentOutcomeDate
10 2015-01-09

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.