[5280655] Complainant reported that patient complained of bad taste in mouth approximately 7 months post-implantation with chin implant. Subsequently, patient complained of pain and discomfort, as well. No biopsy or culture was ever taken, and patient was prescribed two courses of amoxicillin (po) and peridex oral rinse. Device was explanted due to infection approximately one year post-operatively.
Patient Sequence No: 1, Text Type: D, B5


[12875681] Patient: infection, pain. Device: no known device problem. Method: reviewed device history records, sterilization records, environmental monitoring records, and product labeling. Results: device history records review found no assignable cause for the reported events. The sterilization process was within specified parameters, and there have been no other infection-related complaint reports associated with this sterile lot of (b)(4) devices. Product labeling addressed the possibility of infection.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2028924-2015-00002
MDR Report Key4410118
Report Source05
Date Received2015-01-09
Date of Report2015-01-08
Date of Event2014-12-10
Date Mfgr Received2014-12-19
Device Manufacturer Date2013-08-01
Date Added to Maude2015-01-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCRAIG ARTHUR
Manufacturer Street6025 NICOLLE STREET
Manufacturer CityVENTURA CA 93003
Manufacturer CountryUS
Manufacturer Postal93003
Manufacturer Phone8053399415
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCONFORM (TM) EXTENDED ANATOMICAL CHIN
Generic NameCHIN IMPLANT
Product CodeFWP
Date Received2015-01-09
Model NumberNA
Catalog NumberCEAC-L
Lot Number851364
ID NumberNA
Device Expiration Date2018-08-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerIMPLANTECH ASSOCIATES, INC.
Manufacturer AddressVENTURA CA US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2015-01-09