APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR 4407205

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04,06,07 report with the FDA on 2015-01-10 for APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR 4407205 manufactured by Life Technologies Holding Pte. Ltd..

Event Text Entries

[17935252] A customer reported a burning smell coming from the applied biosystems 7500fast dx (cat. No: 4407205, sn:(b)(4)). No burn marks on the instrument were observed. An error message with a flashing red light was present and data was lost during this event. No patient involvement was reported. No patient or user death or serious injury was reported. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[18316268] Implant and explant dates: device is an instrument and is not implanted/explanted. No patient involvement was reported. No patient or user death of serious injury was reported. The instrument was investigated on site. Problem resolved by replacing with the block assembly and the microcontroller board. Event investigation is currently in progress. Device intended use the applied biosystems 7500 fast dx real-time pcr instrument with sds software version 1. 4 is a real-time nucleic acid amplification and detection system that measures nucleic acid signals from reverse transcribed rna and converts them to comparative quantitative readouts using fluorescent detection of dual-labeled hydrolysis probes. The 7500 fast dx real-time pcr instrument is to be used only by technologists trained in laboratory techniques, procedures, and on use of the analyzer.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003673482-2015-00001
MDR Report Key4412179
Report Source04,06,07
Date Received2015-01-10
Date of Report2014-12-10
Date of Event2014-12-10
Date Mfgr Received2014-12-10
Device Manufacturer Date2010-12-01
Date Added to Maude2015-01-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKELLI TANZELLA
Manufacturer Street3175 STALEY ROAD
Manufacturer CityGRAND ISLAND NY 14072
Manufacturer CountryUS
Manufacturer Postal14072
Manufacturer Phone7743122
Manufacturer G1LIFE TECHNOLOGIES HOLDING PTE. LTD.
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAPPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR
Generic NameINSTRUMENTATION FOR CLINICAL MULTIP
Product CodeNSU
Date Received2015-01-10
Catalog Number4407205
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerLIFE TECHNOLOGIES HOLDING PTE. LTD.
Manufacturer AddressSN


Patients

Patient NumberTreatmentOutcomeDate
10 2015-01-10