[15351868] Customer notified that two of their chairs were moving on their own due to a faulty pendant. No pt info was provided as it was a general concern. No adverse event occurred.
Patient Sequence No: 1, Text Type: D, B5


[15641387] The immediate action taken was to replace older style pendants with newly designed pendants for all devices in the facility. There was not an adverse event. It was mentioned that there would be patients on the device. It was presented to tmm as a general concern, not specifically related to any one pt. Transmotion medical inc has designed a new pendant which was released in 05/2013. The new pendant is considerably more durable than the older style pendants. The new pendant has been in production for 17 months and there have been no reports of the new pendant cause the device to move on its own. Additionally, in 11/2014, transmotion medical inc launched a product recall due to previous customer complaint and corresponding mdr data. This mdr is for voluntary reporting purposes only.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004082462-2014-00004
MDR Report Key4417561
Report Source06
Date Received2015-01-08
Date of Report2015-01-08
Date of Event2014-12-09
Date Mfgr Received2014-12-09
Device Manufacturer Date2013-04-01
Date Added to Maude2015-02-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street1441 WOLF CREEK TRAIL
Manufacturer CitySHARON CENTER OH 44274
Manufacturer CountryUS
Manufacturer Postal44274
Manufacturer Phone3302394192
Single Use3
Remedial ActionIN
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRANSMOTION MEDICAL INC
Generic NameTMM4 MULTIO-PURPOSE STRETCHER-CHAIR
Product CodeGBB
Date Received2015-01-08
Model NumberTMM4-XWB
Catalog NumberTMM4-XWB
Lot Number5379-5380
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTRANSMOTION MEDICAL INC.
Manufacturer Address1441 WOLF CREEK TRAIL SHARON CENTER OH 44274 US 44274


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-01-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.