SUPPORT ARM 177

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06,07 report with the FDA on 2014-12-04 for SUPPORT ARM 177 manufactured by Maquet Critical Care Ab.

Event Text Entries

[18790029] Imp ref # (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[18936467] More info surrounding the event has been sought. A supplemental medwatch will be provided when investigation is finished.
Patient Sequence No: 1, Text Type: N, H10


[63281853] Our investigation of the returned support arm has been completed. The investigation showed no indications of low friction in the joints. According to the customer, they had tightened the support arm before returning it. The support arm was stretched out and bent several times without friction reduction or a self-lowering at the level of the joints. The support arm joints were dis-mounted to check the friction rings. The friction rings were not worn out or damaged. No fault could be confirmed from inspection of the friction rings or the joints themselves. According to the installation instructions for the support arm the support arm joints shall be adjusted if necessary. Our conclusion is there is no fault with the support arm, the cause for the reported incident was that the support arm joints were not tightened. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8010042-2014-00549
MDR Report Key4420209
Report Source01,05,06,07
Date Received2014-12-04
Date Mfgr Received2014-11-07
Date Added to Maude2015-01-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional0
Initial Report to FDA0
Report to FDA0
Event Location0
Manufacturer ContactMAGNUS LYNDQVIST
Manufacturer CitySOLNA S-17154
Manufacturer CountrySW
Manufacturer PostalS-171 54
Manufacturer Phone7307705
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSUPPORT ARM 177
Product CodeIOY
Date Received2014-12-04
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMAQUET CRITICAL CARE AB
Manufacturer AddressSOLNA SW


Patients

Patient NumberTreatmentOutcomeDate
10 2014-12-04

© 2026 FDA.report