[5416784]
Percutaneous coronary intervention: successful deployment of 3. 5 x 18mm resolute des to mid-lcx, 75% to 0% with excellent angiographic result and timi 3 flow. Dissection was noted proximal to the mid-stent, successfully covered with stent in proximal lcx. Successful deployment of 3. 5 x 18mm resolute des to proximal lcx, 75% to 0% with excellent angiographic results and timi 3 flow. Patient underwent a complex intervention and had chest pain during the procedure. No complications were noted during the procedure.
Patient Sequence No: 1, Text Type: D, B5
[12923218]
Catheter was returned and was inspected both visually and functionally. There were no visual evidence of failure or mis-use. The catheter had no visual signs of kinking or breakage. Catheter was shown to be fully functionally, tacking well through a simulated arch and producing both an ivus image as well as nir data. No problem found.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 3004722468-2014-00003 |
MDR Report Key | 4431960 |
Report Source | 99 |
Date Received | 2015-01-16 |
Date of Report | 2015-01-15 |
Date of Event | 2014-12-16 |
Date Mfgr Received | 2014-12-17 |
Date Added to Maude | 2015-01-21 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | STEVE CHARTIER |
Manufacturer Street | 34 THIRD AVENUE |
Manufacturer City | BURLINGTON MA 01803 |
Manufacturer Country | US |
Manufacturer Postal | 01803 |
Manufacturer Phone | 7813459593 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | TVC INSIGHT CATHETER |
Generic Name | NIRS IVUS CORONARY IMAGING CATHETER |
Product Code | OGZ |
Date Received | 2015-01-16 |
Returned To Mfg | 2014-12-23 |
Model Number | TVC-C195-22 |
Lot Number | D01234 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | INFRAREDX INC. |
Manufacturer Address | 34 THIRD AVENUE BURLINGTON MA 01803 US 01803 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Life Threatening; 2. Required No Informationntervention | 2015-01-16 |