EPIPHANY

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2015-01-21 for EPIPHANY manufactured by Sybronendo.

Event Text Entries

[5387394] A doctor alleged that two (2) patients had required endodontic re-treatment for root canals that had previously been filled with epiphany. This is the first of two (2) reports.
Patient Sequence No: 1, Text Type: D, B5


[12972411] Specific patient information such as age and weight was not provided. The doctor reported that the initial root canal treatment was performed approximately five (5) years ago by a different endodontist and that epiphany root canal sealant material had been used for the initial procedure. The doctor diagnosed the tooth with a par lesion associated with the palatal root. The doctor cleaned the affected root canals in the patient's tooth, packed it with calcium hydroxide, and prescribed antibiotics for treatment. The doctor reported that black material was found within the root canal during re-treatment and removal of the epiphany material. The patient will return to the office at a later date for completion of the root canal procedure. At this time, the patient is doing fine. Sybronendo has requested that the customer report any new information with regard to this patient. An update will be provided if any new information becomes available. The product involved in the alleged incident was not returned and no lot number was provided; therefore, no further evaluation can be conducted. Upon the investigations of previous comparable incidents, it was found that the black material seen within the product was bismuth sulfide. It is produced by the reaction between bismuth oxychloride in the realseal root canal filler and a protein (most likely from a bodily fluid). The possible causes for the formation of bismuth sulfide are primarily due to technique variations and deviations from the instructions for use of the realseal system, and include overheating and incomplete flushing of naocl from the root canal. Overheating can degrade the realseal canal filler and possibly lead to improper sealing and leaking, which can result in the leeching of proteins into the root canal and subsequent bismuth sulfide formation. Failure to completely flush all naocl from the root canal can also result in the formation of bismuth sulfide. Retreatment of patients' root canals is required to remove bismuth sulfide. This mdr is being submitted because of the medical intervention for retreatment. The previous investigations led to the conclusion that the product itself, when used according to the instructions for use, performs as intended and yields acceptable results.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2016150-2015-00001
MDR Report Key4438981
Report Source05
Date Received2015-01-21
Date of Report2014-12-31
Date Mfgr Received2014-12-31
Date Added to Maude2015-01-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMRS. KERRI CASINO
Manufacturer Street1717 W. COLLINS AVE
Manufacturer CityORANGE CA 92867
Manufacturer CountryUS
Manufacturer Postal92867
Manufacturer Phone7145167634
Manufacturer G1SYBRONENDO
Manufacturer Street1332 SOUTH LONE HILL AVENUE
Manufacturer CityGLENDORA CA 91740
Manufacturer CountryUS
Manufacturer Postal Code91740
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEPIPHANY
Generic NameRESIN, ROOT CANAL FILLING
Product CodeKIF
Date Received2015-01-21
OperatorOTHER
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSYBRONENDO
Manufacturer Address1332 SOUTH LONE HILL AVENUE GLENDORA CA 91740 US 91740


Patients

Patient NumberTreatmentOutcomeDate
101. Other; 2. Required No Informationntervention 2015-01-21