XPZC 13-14 CURETTE XPZC13-14

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06,07 report with the FDA on 2015-01-20 for XPZC 13-14 CURETTE XPZC13-14 manufactured by American Eagle Instruments, Inc..

Event Text Entries

[20016939] I was not able to interview the customer/end-user who was working with the product when they experienced the event. I had very limited info available to me. A 4+ year old dental hand instrument (specifically a curette with a life expectancy of 6-12 months) was being used on a patient's mouth (unsure of the exact procedure being performed) when the point of the instrument (the working portion) came out of the insert (a component of the handle portion of the instrument which joins the point to the handle) and lodged in the patients throat. The loose point was removed resulting in 'minor' injury to the patient. There was no indication of what action or activity took place or was taking place to cause the point to come out of the insert.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1000127001-2014-00001
MDR Report Key4456932
Report Source06,07
Date Received2015-01-20
Date of Report2015-01-16
Date of Event2014-11-30
Date Mfgr Received2014-11-18
Device Manufacturer Date2010-09-02
Date Added to Maude2015-01-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. BRAD DAVIS
Manufacturer Street6575 BUTLER CREEK RD.
Manufacturer CityMISSOULA MT 59808
Manufacturer CountryUS
Manufacturer Postal59808
Manufacturer Phone4065327310
Manufacturer G1AMERICAN EAGLE INSTRUMENTS, INC.
Manufacturer Street6575 BUTLER CREEK RD.
Manufacturer CityMISSOULA MT 59808
Manufacturer CountryUS
Manufacturer Postal Code59808
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameXPZC 13-14 CURETTE
Generic NameUNIVERSAL CURETTE
Product CodeEKT
Date Received2015-01-20
Model NumberXPZC13-14
Catalog NumberXPZC13-14
Lot Number0IB
Device Expiration Date2011-03-30
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerAMERICAN EAGLE INSTRUMENTS, INC.
Manufacturer Address6575 BUTLER CREEK RD. MISSOULA MT 59808 US 59808


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2015-01-20