HUDSON DUAL HEATED DUAL DRAIN 880-36KIT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2015-01-29 for HUDSON DUAL HEATED DUAL DRAIN 880-36KIT manufactured by Teleflex Medical.

Event Text Entries

[5404215] The customer alleges that the concha column was leaking. No pt injury reported.
Patient Sequence No: 1, Text Type: D, B5


[12909913] (b)(4). A visual, functional and dimensional inspection of the product involved in the complaint could not be conducted since the product was not returned. Device history record (dhr) investigation of batch number (b)(4) did not show issues related to complaint. No nonconformity reports originated for lot in question that can be associated to the complaint reported. The product was assembled and inspected according to specifications. No corrective action can be established since the device sample or a picture of it is not available for eval. Customer complaint cannot be confirmed due to the lack of product sample or picture. An attempt to duplicate the failure mode was made but at the time there is no inventory of the involved product code available at the facility nor is it being manufactured at the time. If the physical sample becomes available this complaint will be reopened.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004365956-2015-00052
MDR Report Key4474996
Report Source05,06,07
Date Received2015-01-29
Date of Report2015-01-08
Date of Event2015-01-06
Date Mfgr Received2015-01-08
Device Manufacturer Date2014-10-01
Date Added to Maude2015-02-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARGIE BURTON, RN, RA
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9194334965
Manufacturer G1TELEFLEX MEDICAL
Manufacturer StreetPARQUE INDUSTRIAL FINSA
Manufacturer CityNUEVO LAREDO, TAMAULIPAS
Manufacturer CountryMX
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHUDSON DUAL HEATED DUAL DRAIN
Product CodeCAG
Date Received2015-01-29
Catalog Number880-36KIT
Lot Number74K1402731
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressRESEARCH TRIANGLE PARK NC US


Patients

Patient NumberTreatmentOutcomeDate
10 2015-01-29

© 2024 FDA.report