ICON IMMUNOCONCENTRATION ASSAY FOR GROUP B STREP 4520

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 1996-10-11 for ICON IMMUNOCONCENTRATION ASSAY FOR GROUP B STREP 4520 manufactured by Hybritech Inco..

Event Text Entries

[16271231] A pt was tested using the group b strep assay lot number 690445. The perston reporting the incident believes but is not certain that the sample was obtained just prior to delivery of the infant. A duplicate specimen obtained at the same time was used to streak a culture plate. The result of the assay was negative. The culture result was positive for group b strep. The culture was confirmed with a strep typing kit, mfr unknown. The infant was very sick upon delivery and subsequently died. A blood culture from the infant was performed and the results were negative for group b strep. It was later determined that the infant died from asphyxia. No sample or kit was received from the customer for testing; therefore, the complaint for culture positive. Negative could not be evaluated. Testing was performed at co using an in-house kit of the same lot number.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2022635-1996-00002
MDR Report Key44798
Report Source05
Date Received1996-10-11
Date of Report1996-10-11
Date of Event1996-09-01
Date Mfgr Received1996-09-17
Device Manufacturer Date1996-05-01
Date Added to Maude1996-10-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameICON IMMUNOCONCENTRATION ASSAY FOR GROUP B STREP
Generic NameGBS
Product CodeGTZ
Date Received1996-10-11
Model NumberNA
Catalog Number4520
Lot Number690445
ID NumberNA
Device Expiration Date1997-01-26
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Age*
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key45708
ManufacturerHYBRITECH INCO.
Manufacturer Address8958 TERMAN COURT P.O. BOX 269006 SAN DIEGO CA 921969006 US
Baseline Brand NameICON STREP B
Baseline Generic NameIMMUNOCONCENTRATION ASSAY FOR DETECTION & CONFIRMATION OF GBS FROM VAGINAL & CER
Baseline Model No*
Baseline Catalog No4520
Baseline IDNA
Baseline Device FamilyICON
Baseline Shelf Life ContainedY
Baseline Shelf Life [Months]12
Baseline PMA FlagN
Baseline 510K PMNY
Premarket NotificationK902991
Baseline PreamendmentN
Baseline TransitionalN
510k ExemptN


Patients

Patient NumberTreatmentOutcomeDate
10 1996-10-11

© 2026 FDA.report