MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2015-01-12 for A5 ANESTHESIA DELIVERY SYSTEM manufactured by Mindray Bio-medical Electronics Co. Ltd..
[5503231]
Customer reported an issue with the a5 anesthesia delivery system display, which may have affected anesthesia monitoring. No patient monitoring was reported.
Patient Sequence No: 1, Text Type: D, B5
[12910017]
Company representative evaluated the system. Corrections included replacement of the system's cpu board, reloading system's software, and calibrating. System was tested to factory's specifications.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 2221819-2014-01123 |
MDR Report Key | 4479803 |
Report Source | 05,06,07 |
Date Received | 2015-01-12 |
Date of Report | 2014-11-21 |
Date of Event | 2014-11-21 |
Date Facility Aware | 2014-11-21 |
Report Date | 2014-11-21 |
Date Mfgr Received | 2014-11-21 |
Date Added to Maude | 2015-03-02 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | WISARA SETHACHUTKUL |
Manufacturer Street | 800 MACARTHUR BLVD. |
Manufacturer City | MAHWAH NJ 07430 |
Manufacturer Country | US |
Manufacturer Postal | 07430 |
Manufacturer Phone | 2019958045 |
Manufacturer G1 | MINDRAY BIO-MEDICAL ELECTRONICS CO. LTD. |
Manufacturer Street | KEJI 12TH ROAD SOUTH |
Manufacturer City | NANSHAN, SHENZHEN 518057 |
Manufacturer Country | CH |
Manufacturer Postal Code | 518057 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | A5 ANESTHESIA DELIVERY SYSTEM |
Generic Name | ANESTHESIA DELIVERY SYSTEM |
Product Code | CBQ |
Date Received | 2015-01-12 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MINDRAY BIO-MEDICAL ELECTRONICS CO. LTD. |
Manufacturer Address | NANSHAN, SHENZHEN CH |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2015-01-12 |