FRESENIUS DIALYSIS DELIVERY SYSTEM 2008K *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2003-03-12 for FRESENIUS DIALYSIS DELIVERY SYSTEM 2008K * manufactured by Fresenius Medical Care North America.

MAUDE Entry Details

Report Number2937457-2003-00008
MDR Report Key448561
Report Source05
Date Received2003-03-12
Date of Event2003-02-12
Date Mfgr Received2003-02-12
Device Manufacturer Date2002-01-01
Date Added to Maude2003-03-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactELVY DIZON, BSN, RN, CNN
Manufacturer Street2637 SHADELANDS DR
Manufacturer CityWALNUT CREEK CA 94598
Manufacturer CountryUS
Manufacturer Postal94598
Manufacturer Phone9252950200
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFRESENIUS DIALYSIS DELIVERY SYSTEM
Generic Name*
Product CodeFIL
Date Received2003-03-12
Model Number2008K
Catalog Number*
Lot Number*
ID Number*
Device Availability*
Device Age1 YR
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key437544
ManufacturerFRESENIUS MEDICAL CARE NORTH AMERICA
Manufacturer Address2637 SHADELANDS DR. WALNUT CREEK CA 94598 US
Baseline Brand NameFRESENIUS DIALYSIS DELIVERY SYSTEM
Baseline Generic NameHEMODIALYSIS MACHINE
Baseline Model No2008K
Baseline Catalog NoF002-40
Baseline IDNA
Baseline Device FamilyNA
Baseline Shelf Life [Months]NA
Baseline PMA FlagN
Baseline 510K PMNY
Premarket NotificationK994267
Baseline PreamendmentN
Baseline TransitionalN
510k ExemptN


Patients

Patient NumberTreatmentOutcomeDate
10 2003-03-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.