SENSE SHOULDER COIL 1.5T 4CH 9896-030-13432

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08,distributor report with the FDA on 2015-02-03 for SENSE SHOULDER COIL 1.5T 4CH 9896-030-13432 manufactured by Invivo Corp.

Event Text Entries

[5344103] A female patient was positioned head first supine and scanned with the shoulder coil for an examination of the left shoulder. The patient stated she felt a tingling sensation at the top the left shoulder. After the examination a small bump was observed that developed into a 4cm size blister on top of the left shoulder.
Patient Sequence No: 1, Text Type: D, B5


[12911594] (b)(4). Method, results, conclusions: the investigation is still ongoing on this event. When the investigation is completed a f/u report will be sent to the fda.
Patient Sequence No: 1, Text Type: N, H10


[31807955] (b)(4). No definite root cause for the injury as sustained by the patient could be determined. The most likely root case is a combination of the technician not using any patient ventilation in combination with 8 consecutive scans with high sar values with-out pauses to allow for the patient body to cool down. Product was not returned to the manufacturer, the customer has an agreement with a third party service organization which was unable to return the product to invivo.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1056069-2015-00004
MDR Report Key4487167
Report Source08,DISTRIBUTOR
Date Received2015-02-03
Date of Report2015-01-30
Date of Event2015-01-09
Date Mfgr Received2015-01-26
Date Added to Maude2015-02-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKENNETH REVENNAUGH, DIR
Manufacturer Street3545 S.W. 47TH AVE.
Manufacturer CityGAINESVILLE FL 32608
Manufacturer CountryUS
Manufacturer Postal32608
Manufacturer Phone3523360010
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSENSE SHOULDER COIL 1.5T 4CH
Generic NameCH SHOULDER
Product CodeMOS
Date Received2015-02-03
Model Number9896-030-13432
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerINVIVO CORP
Manufacturer Address3545 S.W. 47TH AVE. GAINESVILLE FL 32608 US 32608


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-02-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.