MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08,distributor report with the FDA on 2015-02-03 for SENSE SHOULDER COIL 1.5T 4CH 9896-030-13432 manufactured by Invivo Corp.
[5344103]
A female patient was positioned head first supine and scanned with the shoulder coil for an examination of the left shoulder. The patient stated she felt a tingling sensation at the top the left shoulder. After the examination a small bump was observed that developed into a 4cm size blister on top of the left shoulder.
Patient Sequence No: 1, Text Type: D, B5
[12911594]
(b)(4). Method, results, conclusions: the investigation is still ongoing on this event. When the investigation is completed a f/u report will be sent to the fda.
Patient Sequence No: 1, Text Type: N, H10
[31807955]
(b)(4). No definite root cause for the injury as sustained by the patient could be determined. The most likely root case is a combination of the technician not using any patient ventilation in combination with 8 consecutive scans with high sar values with-out pauses to allow for the patient body to cool down. Product was not returned to the manufacturer, the customer has an agreement with a third party service organization which was unable to return the product to invivo.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1056069-2015-00004 |
MDR Report Key | 4487167 |
Report Source | 08,DISTRIBUTOR |
Date Received | 2015-02-03 |
Date of Report | 2015-01-30 |
Date of Event | 2015-01-09 |
Date Mfgr Received | 2015-01-26 |
Date Added to Maude | 2015-02-06 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | KENNETH REVENNAUGH, DIR |
Manufacturer Street | 3545 S.W. 47TH AVE. |
Manufacturer City | GAINESVILLE FL 32608 |
Manufacturer Country | US |
Manufacturer Postal | 32608 |
Manufacturer Phone | 3523360010 |
Single Use | 3 |
Remedial Action | RL |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | SENSE SHOULDER COIL 1.5T 4CH |
Generic Name | CH SHOULDER |
Product Code | MOS |
Date Received | 2015-02-03 |
Model Number | 9896-030-13432 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | INVIVO CORP |
Manufacturer Address | 3545 S.W. 47TH AVE. GAINESVILLE FL 32608 US 32608 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2015-02-03 |