[21237391] The case was proceeding well, and there was no fluid deficit (minus the 100ml on floor). The intrauterine tissue removal portion of the case was started and it was noted that the amount of positive deficit (meaning more fluid than less) was noted to climb steadily. The patient was stable and there were no concern from doctor as the fluid was clear with only a hint of blood noted in fluid. The circulator continued to monitor and keep surgeon informed. The intake and output noted via observation and calculation showed there was no additional fluid. At that point, a scale was obtained and all fluids were weighed and it was calculated that there was a 300ml deficit and not an addition of fluid. A work order was put in to have the fluid management system checked and calibrated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number4489968
MDR Report Key4489968
Date Received2014-11-18
Date of Report2014-11-18
Date of Event2014-10-10
Report Date2014-11-18
Date Reported to FDA2014-11-18
Date Reported to Mfgr2015-02-06
Date Added to Maude2015-02-06
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameSTORZ
Generic NameFLUID MANAGEMENT SYSTEM
Product CodeLTA
Date Received2014-11-18
Model Number*
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityY
Device Sequence No1
Device Event Key0
ManufacturerKARL STORZ ENDOSCOPY-AMERICA, INC.
Manufacturer Address2151 E. GRAND AVENUE EL SEGUNDO CA 90245501 US 90245 5017


Patients

Patient NumberTreatmentOutcomeDate
10 2014-11-18