5MMX33CM DISPOSABLE PROBE&VALVE W/DUAL BAG TUBING C6001 100668401

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06,07 report with the FDA on 2015-02-12 for 5MMX33CM DISPOSABLE PROBE&VALVE W/DUAL BAG TUBING C6001 100668401 manufactured by Applied Medical.

Event Text Entries

[5487558] "suction button sticking in open position and locking making it quite dangerous for laparoscopic proc. "patient status: normal.
Patient Sequence No: 1, Text Type: D, B5


[12857056] Investigation summary: the incident product was not returned for evaluation. In the absence of the subject device, it is difficult to determine the root cause of the incident. A review of the manufacturing records for this lot reveals the lot passed all manufacturing and quality inspections. All final assembled handpieces are leak tested at both suction and irrigation valves to ensure the seal functions properly. Through user's interaction with the handpiece, the smoke evacuation feature may have been unintentionally activated. Applied medical continuously seeks to improve the form, function, and ease of use of its products. As part of this continuous process, applied medical has recently re-designed the handpiece which is intended to minimize the potential for this type of incident to occur. This lot was built prior to the implementation of these enhancements. This document represents our initial/final report.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2027111-2015-00030
MDR Report Key4513611
Report Source01,05,06,07
Date Received2015-02-12
Date of Report2015-02-12
Date of Event2014-08-21
Date Mfgr Received2014-08-26
Date Added to Maude2015-02-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer Street22872 AVENIDA EMPRESA
Manufacturer CityRANCHO SANTA MARGARITA CA 92688
Manufacturer CountryUS
Manufacturer Postal92688
Manufacturer Phone949713-823
Manufacturer G1APPLIED MEDICAL
Manufacturer Street22872 AVENIDA EMPRESA
Manufacturer CityRANCHO SANTA MARGARITA CA 92688
Manufacturer CountryUS
Manufacturer Postal Code92688
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name5MMX33CM DISPOSABLE PROBE&VALVE W/DUAL BAG TUBING
Generic NameFHF
Product CodeFHF
Date Received2015-02-12
Model NumberC6001
Catalog Number100668401
Lot Number1203636
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerAPPLIED MEDICAL
Manufacturer Address22872 AVENIDA EMPRESA RANCHO SANTA MARGARITA CA 92688 US 92688


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-02-12

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.