SUPPORT ARM 176 6405976

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,06,07,company representati report with the FDA on 2015-02-11 for SUPPORT ARM 176 6405976 manufactured by Maquet Critical Care Ab.

Event Text Entries

[5411928] It was reported that the support arm had cracks. There was no patient involvement. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[12812457] The part has been requested back for investigation. A supplemental medwatch will be submitted when the investigation is completed.
Patient Sequence No: 1, Text Type: N, H10


[37073680] The broken support arm was received back for investigation, wherein visual inspection of the support arm confirmed the reported cracks. The support arm was manufactured in august of 2002. The position of the break indicates that the support arm had veen overloaded, however, the true cause for the broken support arm/ has/can not be determined from this investigation. The support arm is designed and tested according to the standard en 60601-01.
Patient Sequence No: 1, Text Type: N, H10


[37073681] (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010042-2015-00048
MDR Report Key4524235
Report Source01,06,07,COMPANY REPRESENTATI
Date Received2015-02-11
Date of Report2015-01-23
Date of Event2015-01-23
Date Mfgr Received2015-01-23
Device Manufacturer Date2002-08-01
Date Added to Maude2015-03-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMAGNUS LINDQVIST
Manufacturer Street2 ROENTGENVAGEN
Manufacturer CitySOLNA S-17154
Manufacturer CountrySW
Manufacturer PostalS-17154
Manufacturer Phone7307705
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSUPPORT ARM 176
Product CodeIOY
Date Received2015-02-11
Catalog Number6405976
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMAQUET CRITICAL CARE AB
Manufacturer AddressSOLNA SW


Patients

Patient NumberTreatmentOutcomeDate
10 2015-02-11

© 2026 FDA.report