UMF MEDICAL 4070 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-02-05 for UMF MEDICAL 4070 * manufactured by United Metal Fabricators, Inc..

Event Text Entries

[19091596] The patient was in a room for a visit with the physician (md), and she stood on the height adjustment pedal (patient was "playing" with equipment) that was placed on chassis table platform of exam table. As the body of the exam table came down, it trapped her foot on the "down" pedal in between the chassis platform and the body of the exam table. The "up" height adjustment was non-functional while the "down" pedal was being pressed (involuntarily since her foot was trapped on it). The nurse (rn) was able to pry the pedal from under the patient's foot and engage the "up" pedal to release the patient's right foot. Swelling was noted and a contusion. There was no broken skin. The patient was sent to imaging for an x-ray of her foot. Impression of the x-ray finding: soft tissue of mid-foot. No fracture or dislocation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number4530610
MDR Report Key4530610
Date Received2015-02-05
Date of Report2015-02-05
Date of Event2014-11-12
Report Date2015-02-05
Date Reported to FDA2015-02-05
Date Reported to Mfgr2015-02-20
Date Added to Maude2015-02-20
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameUMF MEDICAL
Generic NameTABLE, POWERED
Product CodeINQ
Date Received2015-02-05
Model Number4070
Catalog Number*
Lot Number*
ID Number*
OperatorPATIENT
Device AvailabilityY
Device Age1 YR
Device Sequence No1
Device Event Key0
ManufacturerUNITED METAL FABRICATORS, INC.
Manufacturer Address1316 EISENHOWER BOULEVARD JOHNSTOWN PA 15904339 US 15904 3393


Patients

Patient NumberTreatmentOutcomeDate
10 2015-02-05