YASARGIL TI PERM MINI-CLIP BAYO 4MM FT727T

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,07 report with the FDA on 2015-02-05 for YASARGIL TI PERM MINI-CLIP BAYO 4MM FT727T manufactured by Aesculap Ag & Co. Kg.

Event Text Entries

[16462073] Country of complaint: (b)(6). The surgeon placed the clip and made a puncture to dome. After placement it was noticed that the tip of the blade was slightly scissored. Clip was left in patient as is.
Patient Sequence No: 1, Text Type: D, B5


[16812744] Us reporting agent notified on 01/30/2015. Manufacturing site evaluation: we received no product, because the clip is still placed inside the patient. It could be possible, that the surgeon, due to an overload situation in the delicate jaw area a scissoring, had created. This product is 100% inspected during the final inspection. The blade alignment is inspected by a visual inspection. So we can exclude a manufacturing error. There are no other cases with "clip was left in patient with scissoring" in our system. Reviewed the manufacturing file of batch 52034072, the quality records as well as the production documents comply with our specifications and do not present any discrepancies. No other similar incidents have been declared concerning this batch.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2916714-2015-00083
MDR Report Key4544163
Report Source01,07
Date Received2015-02-05
Date of Report2015-02-05
Date of Event2014-12-02
Date Mfgr Received2014-12-02
Device Manufacturer Date2014-06-01
Date Added to Maude2015-03-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNICOLE BROYLES
Manufacturer Street615 LAMBERT POINTE DR
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3145515988
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameYASARGIL TI PERM MINI-CLIP BAYO 4MM
Generic NameANEURYSM CLIP
Product CodeHCH
Date Received2015-02-05
Model NumberFT727T
Catalog NumberFT727T
Lot Number52023072
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerAESCULAP AG & CO. KG
Manufacturer AddressTUTTLINGEN DA


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-02-05

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.