* DYNDBARD8

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-02-04 for * DYNDBARD8 manufactured by Medline Industries, Inc..

Event Text Entries

[20979314] During a procedure to start a peripheral intravenous (iv) line in the patient's left wrist, the clinician attempted to connect an extension set from the iv start kit and the connection immediately began to leak. Several attempts made to loosen, disconnect and re-thread luer connection had failed. Two more attempts to use extension sets from additional opened iv start kits still failed to correct the leak. The clinician then opened an individual extension set from a different manufacturer and a tight non-leak connection was made. The clinician decided to remove the iv catheter along with all leaking extension sets and send them to the biomedical engineering department for logging and manufacturer analysis. An addition iv catheter was started along with the second manufacturer set with no additional leaks or issues. There have been additional reported failures or leaks when using the components of the iv start kit. This is an assembled kit with items from various manufacturers. The biomedical engineering department notes that the assembler of kit is the manufacturer of the device (iv start kit). No individual component manufacturer information is provided in the packaging. Manufacturer response: we are attempting resolution on an ongoing issue.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number4544365
MDR Report Key4544365
Date Received2015-02-04
Date of Report2015-02-04
Date of Event2015-01-17
Report Date2015-02-04
Date Reported to FDA2015-02-04
Date Reported to Mfgr2015-02-25
Date Added to Maude2015-02-25
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand Name*
Generic NameIV START KIT
Product CodeLRS
Date Received2015-02-04
Model NumberDYNDBARD8
Catalog NumberDYNDBARD8
Lot Number14IB2246
ID Number*
OperatorNURSE
Device AvailabilityY
Device Age1 DA
Device Sequence No1
Device Event Key0
ManufacturerMEDLINE INDUSTRIES, INC.
Manufacturer AddressONE MEDLINE PLACE MUNDELEIN IL 60060 US 60060


Patients

Patient NumberTreatmentOutcomeDate
10 2015-02-04

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.